Important for D-Tron plus users.
Subject: [IP] Dtron Plus Documentation Error
HEADS UP ANY DISETRONIC EMPLOYEES!!
It's a good thing I only took 5 units last night
(still 20% of my total daily dose!). I had meant to
spread it over a couple of hours since I was grazing,
but hit the check button instead of the menu button
when I wanted to set the time. It all got delivered
at once by mistake and I couldn't stop it following
the directions in the manual.
So, I'm messing around today trying to figure out why
the "menu-menu-check" process doesn't work when I try
to stop a bolus.
I finally figured it out. I have discovered that the
documentation is, at best, mis-leading and, at worst,
WRONG! I've left a message for someone on the
Disetronic support desk to call me back. Some folks
here might be interested in knowing this in the
meantime.
I opened the Dtron+ manual and turned to the page with
directions for stopping an extended bolus. It states
"to stop an extended bolus use menu-menu-check to
place the pump into the stop mode..."
Menu - menu - check DOES NOT WORK for a 0:00 duration.
(I at least remembered to unhook myself before doing
a 10 unit test today).
There is no reference on the stopping an extended
bolus page to a 0 duration having special
circumstances or different procedures. (The extended
bolus stop process is the first page I looked at since
I wanted to stop an extended bolus.)
I finally turned to the "how to set an extended bolus"
documentation (ON A COMPLETELY DIFFERENT PAGE) which
states that when you do an extended bolus with a
duration of 0:00 it "is the same as a "scroll" bolus."
Fine - but still no reference that different
processes apply under that circumstance. Like -
shouldn't it say REFER TO THE SCROLL BOLUS PAGE FOR
FURTHER DIRECTIONS?
FINALLY, I looked at the "scroll" bolus directions and
tried THAT stop procedure - which is to hold down the
^ or v key for 3 seconds. Viola - that stop procedure
magically worked.
The documentation NEEDS TO BE CHANGED!!!!